Flow Reports > Testing of a 10 to 48-inch Insertion Magnetic Flow Meter

 

ITRC Flow Report No. F 20-003 

Testing of a 10 to 48-inch Insertion Magnetic Flow Meter

Download Now Download at Digital Commons

 

With funding from the US Bureau of Reclamation, California – Great Basin Region (Region 10), the Cal Poly Irrigation Training & Research Center (ITRC) conducted flow measurement testing on one (1×) 10 to 48-inch Seametrics EX21x insertion magnetic meter (“Seametrics”, “test meter”) in April 2020. Figure 1 shows the meter installed at the testing facility. The testing determined the accuracy of the test meter in a straight pipe configuration across a range of typical water velocities (0.5 to 5.8 feet per second). Testing was performed at the ITRC Water Resources Facility (WRF) using calibrated measurements from a 30-inch Krohne Enviromag 2100 (“Krohne”, “standard meter”) as the standard flow rate for all tests.

 

Styles, S., J. Evans, S. Layous, and C. Tenbrink. 2020. Testing of a 10 to 48-inch Insertion Magnetic Flow Meter. Irrigation Training & Research Center, California Polytechnic State University, San Luis Obispo. ITRC FLow Report No. F 20-003.


Acrobat Reader is required to open this file.  If Acrobat Reader is not installed in your system, click here for a free download.